Invacare MicroAIR MA65 Alternating Pressure & Low Air Loss Mattress System

Invacare

Order
MSRP: $4,702.21
$2,259.00
(You save $2,443.21 )
Availability:
NO RETURNS ON THIS PRODUCT

Free Shipping
Lowest Price Guarantee

For more details, SEE BELOW.
Order Here
MSRP: $4,702.21
$2,259.00
(You save $2,443.21 )

The Invacare MicroAIR Alternating Pressure Mattress Systems are a powered alternating pressure mattress replacement designed to aid in the prevention and treatment of pressure ulcers (bed sores). The MA65 reduces interface pressure by alternately inflating and deflating 20 individual 8-inch air cell bladders every 5, 10, 15 or 20 minutes. 

Patients with immobility or perceptual problems who remain bedridden for extended periods can develop open sores at points where the weight of the body compresses the skin, often under bony prominences like the hips or shoulders. These sores are called pressure ulcers. 

The alternating pressure mattress replacement system helps in pressure ulcer treatment and management by shifting pressure to different parts of the body in a repeating cycle, which prevents any body area from being under pressure all the time.

  • New wireless fowler technology, real time pressure monitoring and energy conserving power units combine to make MicroAIR therapeutic support surfaces the most efficient and dynamic support surfaces in the market
  • Pressure reduction helps the caregiver to prevent and treat Stage I to Stage IV pressure ulcers and promote capillary circulation
  • Horizontal configuration of 20, 8-inch air bladders and a 2" foam safety pad combine to create a full depth 10" pressure reducing mattress replacement system
  • Fits most standard hospital beds
  • Unique and patented designs of the power unit combine comfort and technology for patient relief and inspire caregiver confidence
  • Control panel is designed with icons to provide easy, intuitive operation for caregivers 
  • 50 LPM compressor has alarm, alarm silence, and lock out features
  • Mattress can be set to alternating or static mode 
  • Alternating mode available in treatment times of 5, 10, 15 or 20 minute cycles
  • Max Inflate mode for patient transfers
  • On-Demand Low Air Loss option offers additional prevention and/or treatment of pressure ulcers
  • Mattress has non-skid bottom and secured to bed with 10 nylon web buckled hold-downs
  • Waterproof, anti-microbial dual quilted top cover with slip layer helps protect against friction and shear
  • Mattress will not bottom out with hoses attached to the compressor power unit, even without power
  • Specially designed connectors are inset to help protect them from accidental dislodging when in use 
  • Built-in 2" foam safety pad for additional protection during power outages if hoses are disconnected from power unit
  • Control unit and mattress are latex-free
  • Mattress width available in 36", 42" or 48" wide
  • Weight capacity is 350 lbs. for standard 36" x 80" mattress
  • Weight capacity is 1000 lbs. for 42" and 48" wide mattresses

CLICK HERE FOR OWNER'S MANUAL

To speak with a customer service representative, call 1-800-956-6616

  • Alternating pressure and low air loss redistribute pressure points to prevent bed sores
  • 50 LPM pump alternately inflates and deflates air cells
  • Low air loss function helps keep skin cool and dry
  • Adjust firmness and cycle time for individual comfort
  • Max inflate mode for patient transfers
  • 8" alternating pressure air cells and 2" foam base
  • Fluid resistant, vapor permeable, low shear quilted cover
  • Multiple sizes available 36", 42" or 48" wide
  • Weight capacity starting at 350 lbs.
Electrical Power: 115VAC, 50/60Hz, 1A
Max Power Consumption: 30 ± 10 W
Circuit Protection: Dual fused, 250 V, 1 A fast blow fuses
Mode of Operation: Continuous
Max Flow Rate: 50 ± 15 LPM
Max Flow Pressure: 35 ± 5 mmHg
Max Flow Timer: 30 minutes
Support Surface Inflation Time: 5 - 10 minutes
Patient Comfort Control Pressures
Soft Pressure: 8 ± 4 mmHg
Firm Pressure: 32 ± 4 mmHg
Cycle Time: 5, 10, 15 or 20 minutes
Operating Conditions
Ambient Temperature: 50°F - 95° F
Relative Humidity: 30% - 75% Non-condensing
Atmospheric Pressure: 70 - 106 kPa
Storage and Shipping Conditions
Ambient Temperature: -40°F - 158° F
Relative Humidity: 10% - 100% Non-condensing
Atmospheric Pressure: 50 -106 kPa
Safety Code Approval: 16 CFR 1632, 16 CFR 1633,
Mattress Size:
Width: 36", 42" or 48"
Length: 80"
Height: 10"
Power Unit Size: 15"W x 11"H x 6.5"D
Power Unit Weight: 15 lbs.
Weight Capacity
36" Wide Mattress: 350 lbs.
42" or 48" Wide Mattress: 1000 lbs.
Cover: Waterproof
Anti-microbial
70 Denier Quilted Nylon
1000 Denier Nylon Bottom
Limited Warranty
Power Unit: 2 years
Mattress: 1 year
Cover: 1 year